توضیحات
اسپکتروفتومتر PG Instruments T110
فروش و خدمات پس از فروش اسپکتروفتومتر PG Instruments T110 در ایران
T110+ Double Monochromator UV-Visible Spectrophotometer
High Performance Dual Monochromator – Double Beam Spectrophotometer with ultra low stray light
The T110+ Series Spectrophotometer incorporates dual monochromator technology making it well suited to even the most demanding of applications in all areas of UV-Visible Spectroscopy including
Pharmaceutical Industry
Material Science
Food Safety
Biotechnology
Research
Wireless instrument control and data acquisition can be achieved from a Table PC using Wi-Fi technology allowing the analyst to move freely around the laboratory whilst also keeping track of sample measurements. The T110+ is fully controlled via a GLP Version of UV-Win Software
The Optical Design for the T110+ gives extremely low stray light characteristics (0.00004% T NaI, 220 nm) which gives the instrument a Large Photometric Range of -6.0 to +6.0Abs units. Measurements at deep ultra-violet wavelengths can also be achieved with use of Nitrogen purged optics
The instrument can be optically configured to suit the needs of the sample by means of a continually adjustable slit for precise control of spectral resolution and beam size adjustment by means of an attenuating wheel. Precise wavelength accuracy is ensured by an integrated Mercury Emission Lamp used for automatic correction of spectral wavelength deviation
A wide range of specialised accessories are available to suit the specific requirement of the sample, these include
Both 60mm and 150mm Integrating Sphere for Diffuse reflectance measurements
Absolute, and Specula reflectance measurements accessories
Polarizing Optics
Thermostatic Cell Holders for temperature control
Various long and short path length cell holders
Automated cell changers for both sample and reference beams
Tablet dissolution accessory for pharmaceutical Quality Control
DETAILED SPECIFICATION
Optical System Dual Monochromator – Double Beam Optics
D2 Lamp – UV Region
Lamp – Visible Region
Hg Lamp – Wavelength Calibration
Wavelength Range 185 – 900nm
Wavelength Accuracy ±0.2nm
Wavelength Reproducibility ≤ 0.1nm (D2 lamp)
Spectral Bandwidth 0.1 –5.0nm Continually Adjustable
Stray Light
≤0.0001%T (NaI , 220nm)
≤0.00004%T ((NaNO3, 360 nm)
Photometric Accuracy
±0.004A @2.0 A
±0.003A @1.0A
±0.002A @0.5 A
±0.3%
Photometric Reproducibility
≤0.002A @2.0 A
≤0.0008A / 1.0A
≤0.0004A/ 0.5A
≤0.1%
Baseline Flatness ±0.0008Abs
Noise
0%T Noise:=0.01%
100% T Noise:=0.1%
Communication Port USB or RS232C or WiFi